Feature:
45 wide range live SLO imaging.
Ultra fine quality retinal imaging using averaging technique(up to 50 images).
SLO based real-time retinal tracking effective lyreduces artifacts caused by eye movement.
For comprehensive glaucoma analysis, OSE 4000 offers two scan patterns, glaucoma cube scan in macular area and glaucomacube scanin discarea Evenly distributed sampling point with 200 * 200 A-scans provides reliabe information for early glaucoma detection and management.
Optical Coherence Tomography Angiography (OCTA) is a new non-invasive imaging technique that allows the detailed study of low within the vascular structure of the eye without the need of dye injections.
HD OCT imaging system at 80000 A-scans/s.
Technical Data | |
OCT Imaging | |
Methodology | Spectral domain OCT |
Optical source | Super luminescent diode (SLD), 840 nm |
Scan speed | 80,000 A-scans/s |
Axial resolution (optical) | 5 microns (optical), 3.6 microns (digital) |
Transverse resolution | 15 microns (optical), 3 microns (digital) |
A-scan depth | 3mm |
Diopter range | -20 to+ 20 diopters |
Scan patterns | Macular: HD line scan (6 or 12 mm), 3D scan (6 mm x 6 mm), 6-line radial scan, Multi (X-Y: 5 x 5) |
Disc: 3D scan (6 mm x 6 mm) | |
Anterior: HD line scan (6 or 16mm), 6-line radial scan | |
Fundus Imaging | |
Methodology | Line scanning laser ophthalmoscopy (LSLO) |
Minimum pupil diameter | 3.0mm |
Field of view | 45 degrees |
VASCANTM OCTA Module | |
VASCAN advance | |
Scanning volume/area | 3mm x 3mm 256 x 256 A-scans |
6mm x 6mm 360 x 360 A-scans | |
8mm x 8mm 360 x 360 A-scans | |
12mm x 8 mm 540 x 360 A-scans | |
Algorithm | C-OMAG |
Segmentation options | Encoded, Vitreousretina Intrerface(VRI), Superfcial retina, Deepfcial retinal, Avascular, |
Choriocapillaris, Choriod, Custom | |
Quantitative analysis | Yes |
Weight | 30.5 kg |
Dimension | 532 mm (L) x 360 mm (W) x 540 mm (H) |
Source voltage | AC 100 – 240 V, 50 Hz – 60 Hz |
Power input | 90VA |
Send Inquiry
Please fill out the form below, and write down your requirements. The field marked with red * is required.